Wind Speed
- Flow field specification
- Digital micromanometer
- Wind direction measurement control
First test section | |
Cross-sectional dimension | 0.8m×0.8m×1.6m |
Flow velocity range | 0.4~70 m/s |
Flow field homogeneity | ≤1%(10m/s~50m/s) |
Flow field stability | ≤0.5%(10 m/s~50 m/s);≤1%(≥50 m/s) |
Airflow drift angle | vertical direction,horizontal direction ≤1° |
Airflow velocity | 0.5% |
Axial static pressure gradient | ≤0.005 |
Second test section(Wind speed @10 m/s) | |
Cross-sectional dimension | 1.3m×1.3m×2.5m |
Flow velocity range | 0.2~27m/s |
Flow field homogeneity | ≤1% |
Flow field stability | ≤0.5% |
Airflow drift angle | vertical direction,horizontal direction ≤1° |
Airflow velocity | / |
Axial static pressure gradient | ≤0.005 |
Specification | |
Working environment | Temperature 0~40℃, Humidity 5~90% RH,barometric pressure 500~1060 hPa |
System dimension | Wind tunnel:18.75×8.05×2.75m(Without holder),control console:2.80×0.92×1.25m |
Specification | |
Measuring range | 0~7500Pa |
Resolution | 0.01Pa |
Differential pressure measuring accuracy | 0.01%FS |
Differential pressure calibration validity | 1 year |
Specification | |
Dial resolution | 1° |
Dial circular rotation range | -360°,+360° |
Dial circular rotation angle control accuracy | ≤1° |
Dial circular rotation control resolution | ≤0.2° |
Zero-positioning device position error | ≤0.5° |
- Products