DST1200
CO₂ Transmitter
Specifications
· NDIR (non-dispersive infrared) waveguide technology
· CO₂ measurement accuracy from 40ppm to 70ppm optional
· Varies measurement range available
· Optional analog signal or RS485 output
· Wide angle view LCD screen with CO₂ readings in real time
· Toxicity and halogen free material, complying with GXP standard
Certificate
Specification
Measuring Data
DP Measuring Range |
0~2000ppm,0~5000ppm,0~10000ppm |
DP Accuracy |
±70ppm,±40ppm |
DP Resolution |
1ppm |
General
Power Supply |
4~20mA (0.02×RL+10V)<UV<30V DC 5V~28V DC (0~5V 8V DC~28V DC 0~10V 14V DC~28V DC) |
Working Temperature |
-30℃~70℃ , -40℃~85℃(No display) , 0℃~50℃(Best fit) |
Storage Temperature |
-50℃~90℃ |
Transmitter Type |
3 wire 4~20mA,Voltage output,RS485 |
Analog Output |
4~20mA RL<(UV-10) / 0.02< 500Ω , 0~10V -1mA<IL<1mA , 0~5V -1mA<IL<1mA 0~1V -1mA<IL<1mA |
Power Consumption |
Average currents 35mA , peak current 305mA(24V DC) |
Digital Output |
RS485(Only for DD version) |
ESD Protection |
±25KV |
IP Class |
IP50 |
Sensing Probe
General Feature
Dimensions |
100.0×100.0×25.5mm |
Housing |
ABS+PC,flame-retardant,toxicity and halogen free |
Weight |
83g |
Response Time |
|
Dimensions(Unit : mm): |
Wall Dimensions(Unit : mm) |
|
|