Specifications
· High accuracy with measuring range customized
· Optional analog signal and digital signal
· Single-point and multi-point calibration
· Compatible with Rotronic and Vaisala's probe command
· Optional interface M12 , RJ45 and 8 wire
· Toxicity and halogen free material , complying with GXP
Certificate
Specification
Technical Specification
Temperature measuring range |
-40℃~70℃ |
Temperature measuring accuracy |
±0.3℃ over- 20℃~+60℃, ±0.4℃ over -40℃~+70℃ |
Reported resolution |
0.001℃ |
Repeatability |
0.04℃ (3-sigma value of 25 measurements at constant temperature) |
Long-term drift |
<±0.003℃ per year |
Response time (with filter) |
120 s (63% response in air moving at 1 m/s) |
Humidity measuring range |
0%~100%RH |
Humidity accuracy |
(at 25℃)±1.8%RH over 0 to 80%RH, ±3%RH over 80 to 100%RH,<±1% RH (at other temperatures over -40 to +60°C) |
Short term hysteresis |
<1% RH |
Temperature dependence |
Compensated to<±2% over -20~60℃ |
Typical long-term stability |
<±0.5% per year |
Reported resolution |
0.001% RH |
Temperature sensor |
PT100 1/3DIN |
Humidity sensor |
Humidity capacitance |
Supply voltage |
5V~28VDC(0~5V8V DC~28V DC 0~10V 14V DC~28V DC) |
Output |
Analog signal: voltage 0~1V/0~2.5V/0~5V/0~10V,temperature 4-wire resistance output,digital sianal:RS485 |
Interfaces |
8-wire,M12,RJ45 |
Storage range |
-60℃~+90℃ |
ESD Protection |
+25KV |
IP class |
IP64 |
Housing |
ABS+PC, flame-retardant, toxicity and halogen free |
Dimensions |
Ф18.6x140mm |
Lead wire |
2m/3m/5m (customized) |
Weight |
26g |
DSP1200 Family |
|
Probe Dimensions(Unit : mm): |
|